KOEN中HI

Hardware Architecture

ECG input stage · patient isolation · protection · power · board layout — based on the 2026-10-11 development-start edition (rev3), block-level summary

This page condenses the design documents to the block level. Resistor and capacitor values, connection tables and parts lists are not included. Each item's status is marked as follows.

Confirmed Confirmed as specification Design default Chosen with stated rationale, but not confirmed Unverified Not yet verified against datasheets or original sources

1. Starting point — confirmed specifications

This device touches the human body. Patient-contact isolation and leakage current were therefore taken as the starting point of the design. Development priorities are ① reliability and patient protection → ② new technology and advancement → ③ after-sales service, maintenance and future upgrades, in that order.

ItemDetailsStatus
Intended useHome use · senior-care industry · personal monitoring, with Holter/trend recording. Not an ICU, general-ward or surgical monitorConfirmed
ElectrodesBasic 5 electrodes (RA · LA · LL · RL + 1 chest). The chest electrode is selected as V1 / V5 by a switch and shown on screenConfirmed
Option12-lead, accepting all six C1–C6Confirmed
lead IIIMeasured directly, not calculated (II−I)Confirmed
ECG chipTI AFE159RP8 — on condition that production status is verified with the detailed datasheetConfirmed
SpO2TI AFE4490 in-house circuit, transmissive clip probe. Numbers are "reference value · not calibrated"Confirmed
ProcessorNXP i.MX 95Confirmed
Pulse resistor locationOn the board, not in the lead wiresConfirmed
Power24 hours of continuous operation on battery alone. Two batteries, no interruption during replacement (hot-swap). AC power always connectedConfirmed
Cooling · noiseFanless by default. No audible noiseConfirmed
Measurement while chargingTwo-stage release — interlock kept → leakage and noise during charging pass measurement → releasedConfirmed
SimulatorExternal battery-powered, 1 V p-p output, 1000:1 divider in the main unit, sim ↔ signal switching at the input stageConfirmed
Respiration · NIBPBoth +option. Respiration uses the ECG chip's built-in function ("reference value"); NIBP has a module slot preparedConfirmed
Safety MCUTI RM44L920. Focus on building first (time of adoption TBD)Confirmed

2. Overall block diagram

Patient island (floating ground · patient contact) Isolation barrier · 8 mm copper-free + slot Secondary (non-patient side) Electrode sockets DIN 42802 RA LA LL RL + C (V1/V5) C1~C6 (option cover) 3-stage input protection ① Pulse resistor (board) ② Gas discharge tube ③ Clamp · RC Same for every electrode sim ↔ signal Latching relay SIM input 1000:1 divider AFE159RP8 U1 basic AFE159RP8 U2 12-lead option SpO2 clip probe → AFE4490 Separate section · probe treated as applied part RLD drive · WCT · guard Lead-off · pace detection (on-chip) Island power — UCC12050 secondary → low-noise LDO AFE159RP8 analog · AFE4490 RX/TX (LED) · isolator island side Metal enclosure not connected to island ground Enclosure–island: at least 1 MOPP (target 2 MOPP) ISO7762 ECG ×2 ISO7762 SpO2 ×1 UCC12050 Isolated power NXP i.MX 95 M7 real-time core: acquisition Samples received via isolated SPI Display · alarms · recording V1 / V5 slide switch Read on 2 GPIO lines · does not cross isolation 10.1″-class panel · touch Backlight current dimming GbE 2ch · Wi-Fi/BT USB offline · OTA Main power section Adapter · 2 battery packs · buck

Green = patient island, yellow band = isolation barrier, purple = secondary side, orange = power. The orange dashed chip is the 12-lead option. The only parts crossing the barrier are the isolator (ISO7762) and isolated power (UCC12050) packages.

3. ECG input stage

Chip configuration — two AFE159RP8 Design default

U1 alone handles the basic 5 electrodes. U1 covers the I · II · III (direct) · V (= C − WCT) · calib channels plus pace detection, RLD drive and respiration (option). For the 12-lead option, U2 handles V1–V6, a separate calib, and a II channel for cross-checking between chips.

The reason for two chips is channel count. Doing 12-lead with one 8-channel chip fills it with I · II · V1–V6 alone, leaving no room for the confirmed-spec direct lead III measurement and calib.

12-lead option expansion method Design default

V1 / V5 switch Confirmed (selection · display) · implementation Design default

Connectors · shielding

Lead-off · pace

4. Input protection

Every signal electrode has the same 3-stage protection chain. Per the confirmed spec, the pulse resistors are on the board, not in the lead wires.

① Pulse resistorAbsorbs high-voltage pulse energy, first ESD defense. Split into two pieces to lower the voltage across each
→
② Gas discharge tubeClamps the middle node to island ground to protect later stages
→
③ Current limit + clampResistor limits current; low-leakage diodes hold the signal within the chip supply range
→
RC → AFESeparates the external clamp from the chip's internal ESD diodes, EMI band limiting

5. Patient isolation

Patient islandElectrodes · protection · AFE159RP8 · AFE4490 · island LDO. Floating-ground region, a single island ground plane
⇄
Isolation barrier8 mm copper-free on all layers + slot
Crossing: ISO7762DW (ECG ×2, SpO2 ×1), UCC12050
⇄
Secondary sidei.MX 95 board, main power section, display, communications

6. Power architecture

Medical adapter24 V, 2×MOPP
(GSM90B24)
→
eFuseTPS26630
Reverse polarity · overvoltage · overcurrent
→
Power path · chargingLTC1760
AC priority, both packs discharge simultaneously on outage, charges both packs
→
2 bucksLT8640S ×2
5 V · 3.3 V
Battery packs A · BRRC2054-2 (4S) ×2, hot-swap Design default
Loadsi.MX 95 SoM · panel · LP8866S backlight · peripherals
Patient island powerUCC12050 → in-island LT3042 ×4 + TPS7A2030
Always-on powerTPS70933 · LTC2955 (button · auto power-on on AC return) · interlock logic

Power protection — concept by stage

Eliminating audible noise Confirmed (goal) · method Design default

24-hour battery · fanless

Biggest warning — recalculating patient island power including the isolators' and LDOs' own currents, average power in outage mode comes very close to the 24-hour limit of an end-of-life pack (calculated margin 2–3 %). Therefore the 24-hour measurement is done first.

7. Board layout

ZoneMain blocks
Patient islandElectrode sockets · 3-stage input protection · sim↔signal relay · AFE159RP8 ×2 (U2 is the 12-lead option) · SpO2 AFE4490 (separate section) · RLD/WCT/guard · island LDO
Isolation barrierISO7762DW ×3 (ECG 2, SpO2 1) · UCC12050 · 8 mm copper-free + slot
ProcessorNXP i.MX 95 SoM — M7 real-time core acquires samples, SoM handles display, alarms, recording · safety MCU RM44L920 footprint (time of adoption TBD)
Display10.1″-class panel · capacitive touch · LP8866S backlight (current dimming)
Communications · updatesGbE 2ch · Wi-Fi/BT (built into SoM) · USB offline + internet OTA
Power sectionMedical adapter · eFuse · LTC1760 · 2 battery packs (hot-swap) · LT8640S ×2 · always-on power · interlock
Option slots12-lead (C1~C6) · Respiration (built into ECG chip) · NIBP module

8. Design history

After the first hardware design on 2026-09-24, the main parts changed with the 2026-10-11 specification upgrade decision.

ItemFirst design (2026-09-24) PreviousDevelopment-start edition (2026-10-11)
ECG chipTI ADS1298TI AFE159RP8 ×2 Confirmed (chip)
Electrodes4 (RA · LA · LL + RL)Basic 5 electrodes + V1/V5 switch, 12-lead option Confirmed
Defibrillation protectionNot designed (assumed "do not use simultaneously" labeling)3-stage protection included Design default
ProcessorConsidered i.MX 8M Plus or RK3588 + separate acquisition MCUNXP i.MX 95 (acquisition on M7 real-time core) Confirmed
Display10.5″ FHD~2K considered10.1″-class Confirmed
Battery18650 2S-type battery floating supply2 packs hot-swap, 24-hour target Confirmed · 4S pack Design default
Carried over unchangedPatient island + UCC12050 · ISO7762DW isolation, 8 mm copper-free barrier, switcher synchronization with spread spectrum off, AFE4490 SpO2, external simulator 1000:1 divider input

9. Remaining unverified items · next measurements